How do I use Oakton Ecotestr?
How do I use Oakton Ecotestr?
Calibration Steps
- Start with a pH 7 buffer.
- Remove the cap, turn on the meter, then press the “Cal” button.
- Place the ph meter into the buffer.
- Wait for the meter to stabilize.
- Once the reading is stablized, press the “Cal” button again to lock in the value.
- Repeat these steps for both pH 4 and pH 10 buffers.
How to use oakton pH meter?
Calibration Steps
- Rinse your pH electrode.
- Press the on/off button to switch the unit on.
- Place the electrode in pH 7 buffer solution.
- Press the “Cal” key to put it into calibration mode.
- Wait for the reading to stabilize.
- Press the “Hold/Ent” key to confirm the first calibration point.
- Rinse the electrode.
How do I reset my pH tester?
How to calibrate a pH meter
- Examine the pH electrode. First, check the pH electrode for contamination or damage.
- Flush the pH sensor. Next, flush your sensor with distilled water.
- Immerse the pH electrode.
- Calibrate the pH meter.
- Rinse the pH sensor and repeat.
How to change battery in ecotestr pH 2?
Operating Instructions EcoTestr pH 2 Waterproof Pocket Tester. Changing Batteries: To remove batteries, lift up front battery cover and hold in position before lifting two sides of the pocket clip (Figure 1). Remove old batteries and replace with fresh ones. Note polarity as shown in battery compartment.
How to calibrate the Oakton ecotestr pH 2 Pocket pH meter?
Place the ph meter into the buffer. Make sure the bulb is completely submerged. Wait for the meter to stabilize. This may take up to fifteen seconds Once the reading is stablized, press the “Cal” button again to lock in the value. To end the calibration for any reason, press the cal escape button. Now you are ready to test your solution!
Do you need ATC on ecotestr pH 2?
If your application requires pH measurements that vary greatly, you should use EcoTestr pH 2, which offers 3-point calibration. Do I need Automatic Temperature Compensation (ATC)?
How to setup ecotestr PH2 + waterproof pocket tester?
Press button to enter the Setup Menu and ‘bUFF’ will appear. menu EcoTestr pH2+ Waterproof Pocket Tester Measurement: 1. Remove cap and press button to turn on the tester. 2. Dip sensor in at least 20 mm of test solution. 3. Stir gently and wait for the flashing to stop. 4.